Zkuste hledat podle klíčových slov / objednávky, sériového čísla, např. „CM442“ nebo „technické informace“
Pro zahájení vyhledávání zadejte alespoň dva znaky.
Product picture SS500e single channel H2O, gas analyzer, right angle view

SS500E

SS500E
Please Wait (spinning wheel)
Zjišťování ceny
Cena je dočasně nedostupná
Price on request
aktivace ceny
cena: poskytnout informace
from CHF ??.-
Filtrovat úlohy měření / Aplikace
Měřicí / aplikační úloha

Nakonfigurujte stahování CAD výkresů pro tento produkt

  • Měřicí princip

    TDLAS

  • Product Headline

    An economical gas analyzer for moisture (H2O) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Includes a heated enclosure for the sample conditioning system. Option of a simple sample conditioning system to reduce complexity of the analyzer.

  • Channels

    1

  • Analyte and Measurement ranges

    H2O (Moisture):
    5 to 422 ppmv (minimum)
    5 to 2110 ppmv (maximum)

  • Measured Variables

    Concentration
    Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 700 to 1400 mbara or 700 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    100 to 240 VAC, 50 to 60 HZ
    or
    18 to 24 VDC - optional
    1 Amp maximum at 120 VAC , 1.6 Amps at 24 VDC (unheated)
    2 Amps maximum at 120 VAC (heated)

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C - standard, RS485 and Ethernet - optional
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 2, General Fault and Concentration/Assignable Alarm

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    Type 4X